6060 Spine Road
Boulder, CO 80301 USA
T 303 444 2009
F 303 444 9543
GE Analytical Instruments, Inc. ©2005H Page 1 of 16 DIN 30001 Rev. E
TOC,FLOW SWITCH UPGRADE 2.12CBI
Sievers 800 TOC Analyzer Flow Switch Upgrade
(Flow Switch Kit and Firmware 2.12 CBI)
1. Purpose: To describe the mechanical installation for the flow switch, firmware 2.12 CBI
upgrade installation and operation of the flow switch.
2. Scope: This procedure describes the activities required to install the flow switch and
firmware 2.12 CBI in all Model 800 series TOC analyzers. For TURBO analyzers,
this procedure describes the installation and configuration of the flow switch.
Turbo Analyzers will omit the firmware installation section.
3. Materials and Tools:
7/16" open end or adjustable wrench
1/2" open end wrench
Flat head screwdriver
Phillips head screwdriver
Scriber
Teflon
®
tape
WARNINGS:
A 60-micron filter must be used to ensure proper operation of the flow switch.
Do not over tighten any fitting.
Switch orientation should be no more than 15° from vertical alignment.
Do not install units near ferrous metals, magnets or electromagnetic fields.
Please review the enclosed Operation and Maintenance Manual. There have
been features added to the analyzers. These additions are dependent upon
the firmware version and software version currently in use.
Vue de la page 0
1 2 3 4 5 6 ... 15 16

Résumé du contenu

Page 1

6060 Spine Road Boulder, CO 80301 USA T 303 444 2009 F 303 444 9543 GE Analytical Instruments, Inc. ©2005H Page 1 of 16 DIN 30001 Rev.

Page 2

GE Analytical Instruments, Inc. ©2005H Page 10 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI to save digits. After resetting the clock,

Page 3

GE Analytical Instruments, Inc. ©2005H Page 11 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI Constants Record CONSTANT VALUE NAME CONST

Page 4 - -1.3250000

GE Analytical Instruments, Inc. ©2005H Page 12 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI CONSTANT VALUE NAME CONSTANT VALUE AFTER UP

Page 5 - FRONT OF TOC

GE Analytical Instruments, Inc. ©2005H Page 13 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI CONSTANT VALUE NAME CONSTANT VALUE AFTER UP

Page 6

GE Analytical Instruments, Inc. ©2005H Page 14 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI 6. Firmware Configuration It is necessary

Page 7

GE Analytical Instruments, Inc. ©2005H Page 15 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI ENABLE/DISABLESET POLARITYENTERCLEAR FIGURE

Page 8 - CAUTION:

GE Analytical Instruments, Inc. ©2005H Page 16 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI normally open or closed. Press Clear twice

Page 9 - EXIT? ENT = Y CLR = N

GE Analytical Instruments, Inc. ©2005H Page 2 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI FIGURE ONE: Flow Switch

Page 10

GE Analytical Instruments, Inc. ©2005H Page 3 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI 60 MICRON FILTERFLOW SWITCHREGULATORSAMPLING

Page 11 - Constants Record

GE Analytical Instruments, Inc. ©2005H Page 4 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI TC ZERO CORRECTION-1.3250000ENTERCLEARENT =

Page 12

GE Analytical Instruments, Inc. ©2005H Page 5 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI 5.2.9. Select each consumable, press Enter.

Page 13

GE Analytical Instruments, Inc. ©2005H Page 6 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI 5.3.4. Replace the EPROM from a Zero Insert

Page 14

GE Analytical Instruments, Inc. ©2005H Page 7 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI carrier, gently press the pins of the chip t

Page 15 - SET POLARITY

GE Analytical Instruments, Inc. ©2005H Page 8 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI CAUTION: IF THERE ARE ANY VARIATIONS, CONSTA

Page 16

GE Analytical Instruments, Inc. ©2005H Page 9 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI FIGURE NINE: First Constant Screen 5.6.1.8.