April 2004
4-18 Programming Site Data
Entering the Standard Volume Parameters
(cont.)
2. The next prompt asks for the base pressure.
a. Press
[ENTER] to enter the text box.
b. Use the numeric keys to enter the known pressure.
c. Press
[ENTER] to confirm the entry.
3. The next prompt asks for the atmospheric pressure.
a. Press
[ENTER] to enter the text box.
b. Use the numeric keys to enter the local atmospheric
pressure.
c. Press
[ENTER] to confirm the entry.
4. The final prompt in this window asks for the base temperature
(in degrees F or C).
a. Press
[ENTER] to enter the text box.
b. Use the numeric keys to enter the known temperature.
c. Press
[ENTER] to confirm the entry.
Pressing the [
S] key returns the meter to the Standard Volume
prompt, and then to the Settings tab.
To confirm the entries and return to Operate mode, press [F3]
(OK).
To leave the window without confirming the entries and return
to Operate mode, press
[F2] (Cancel) or the [ESC] key.