Programming Site Data 4-19
April 2004
Entering Inputs in the Standard Volume Option
To enter input parameters in the Standard Volume/Mass Flow
option, return to the Settings tab at the top of the Standard
Volume window. Press the [X] arrow key and
[ENTER] to open
the Inputs window. The screen appears similar to Figure 4-8
below. To step through each parameter, press the [T] key.
Figure 4-8: The Inputs Tab in the
Standard Volume/Mass Flow Option
1. The first prompt asks if the pressure is fixed or active.
a. Use the [
W ] and [X] keys to scroll to the appropriate
radio button.
b. Press
[ENTER] to confirm your selection.
Settings Inputs
Cancell OK
Pressure
Fixed
Active
Bars
Input #
1.01
A
Temperature
Fixed
Active
°C
Input #
15.5
B
Pressure
Temp
Std Vol/Mass Flow
Mass Flow