GE PT878GC Manuel d'utilisateur Page 89

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 308
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 88
April 2004
4-28 Programming Site Data
Entering Skan/Measure Integrate Parameters
To enter parameters for the Skan/Measure Integrate mode in the
Skan/Measure option, press the [X] arrow key once to reach the
Integrate tab. Then press
[ENTER] to open the Integrate window,
which appears similar to Figure 4-12 below.
Figure 4-12: The Integrate Window in the skan/
Measure Option
1. The first prompt, A Divisor, is used in the calculation of the
Measure Mode integrated threshold level and is not normally
changed. The default value for this parameter is 2.5, and
values from 0.1 to 10 are acceptable.
a. Press
[ENTER] to enter the text box.
b. Use the numeric keys to enter the recommended value.
c. Press
[ENTER] to confirm the entry.
2. Next, enter the Noise Level ( -Offset % of Peak) from 0 to
95%, and press
[ENT]. (The default value is 20%.)
a. Press
[ENTER] to enter the text box.
b. Use the numeric keys to enter the recommended value.
c. Press
[ENTER] to confirm the entry.
Integrate
Skan/Measure
Cancell OK
All
A Divisor
Noise Level
Count
Correlate
2.5
Measure Only
5
%
Vue de la page 88
1 2 ... 84 85 86 87 88 89 90 91 92 93 94 ... 307 308

Commentaires sur ces manuels

Pas de commentaire