ValVue Digital Communications Software Ver. 2.80
106
GE Oil & Gas
Dragging Data Points
Move points by dragging them to their new position (Figure 78). The characterization curve
must be non-decreasing and the program does not allow a point to be added or moved to a
position that would create a decreasing segment. To drag a data point:
1. Select a data point by dragging and holding with a left click.
2. Drag the data point to the new location.
Figure 78 Changing Curve by Dragging
Saving Custom Characterization Curves
To save the custom curve:
1. Click OK.
Figure 79 appears.
Figure 79 Save Custom Characterization
2. Click OK to save the custom characterization data.
3. Return to the Calibration screen and click Apply to save the changes.