ValVue Digital Communications Software Ver. 2.80
370
GE Oil & Gas
SG Meter Calibration
Use the Specific Gravity (SG Meter) Calibration function on the Advanced Setup screen to
directly read the specific gravity of the process liquid from the 12400. SG Meter is used to
perform on site new calibration or simulation, with or without liquid.
To complete the SG Meter Calibration function you must first Zero the SG and then perform a
Span reading to arrive at the new SG Calibration.
SG Meter Calibration Zero
To Zero the SG Meter Calibration:
1. Click ZERO.
ValVue 12400 issues the dialog shown in Figure 167.
2. Empty the displacer chamber.
3. Click OK.
Figure 167 Zero SG Meter Dialog
After the Zero reading is complete, a dialog appears (Figure 168).
4. Click OK to close the dialog and return to ValVue 12400.
Figure 168 Zero Completed Message